EasyManua.ls Logo

ABB AWT420 - Ph and Conductivity Connections; Ph;Orp;Pion Sensor Module Connections

ABB AWT420
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AWT420 | UNIVERSAL 4-WIRE, DUAL-INPUT TRANSMITTER | CI/AWT420-EN REV. B
13
pH and conductivity connections
pH/ORP/pIon sensor module connections
NOTICE
ORP (Redox) and Antimony pH sensors do not feature
temperature compensation therefore do not have
temperature sensors or related wiring.
Standard sensors without diagnostic functions
NOTICE
Ensure sensor diagnostics are Off when using standard
sensors without diagnostic functions.
Sensor type RTD wiring
SENSE
1
GUARD
2
REF
3
S.GND
4
RTD 1
5
RTD 2
6
SHIELD
7
RTD 3
8
  Clear Black Red White
  Blue Black Red White

 Clear Black
Red
Red
White
 Clear Black White Red Red

 Blue Black Red White
 Blue Black Red White Grey



 Blue Black Red White Yellow
 Blue Black Red White Yellow Grey

Standard sensors with diagnostic functions
NOTICE
Ensure sensor diagnostics are On when using standard
sensors with diagnostic functions.
Sensor type RTD wiring
SENSE
1
GUARD
2
REF
3
S.GND
4
RTD 1
5
RTD 2
6
SHIELD
7
RTD 3
8
  Blue Yellow Black Green Red White 

 Clear Red Blue Green/Yellow Red White
 Clear Red Blue Green/Yellow Red White Grey


Table of Contents

Other manuals for ABB AWT420

Related product manuals