Electromagnetic Flowmeter
COPA-XE / MAG-XE
4.13 Submenu Alarm
Entry from table
The functions incorporated in this menu can be selected using
the STEP and DATA keys after ENTER is pressed.
Error Register (4.13.1)
MAX-Alarm (4.13.2)
MIN-Alarm (4.13.3)
4.13.1 Error Register
All detected errors (Error 0 to 9, A to C) are stored in this
register. All detected errors will remained stored until the
register is manually reset (using the ENTER key).
In the example, Error 0 (Empty Pipe) and Error 3 (Flowrate >
130 %) have been registered since the last reset.
The following display appears after pressing ENTER:
After pressing ENTER the error register is cleared.
Press the STEP key to access the help text which displays a
clear text description for each error code.
Error 0 = Empty pipe
Error 8 = negative reference voltage too large
Exit the help text information by pressing C/CE.
4.13.2 Setting the MAX-Alarm
The limits for the desired MAX-Alarm value can be set in 1 %
steps from 0 % to 130 %. This value applies to both the forward
and reverse flow directions.
When MAX-Alarm is selected the contact is actuated when the
flowrate exceeds the limit value set. If this occurs, an upward
pointing blinking arrow is displayed.
MAX-Alarm limit value = 110 %
For flowrates >110 % a upward pointing blinking arrow is
displayed next to the flow direction indicator in the 1st line. If the
MAX-Alarm value is set to 0 %, the option is turned off and no
alarm arrow is displayed when the flowrate exceeds the value.
Submenu
Alarm
Error Register
0...3...
Reset: ENTER
Help Text: STEP
Error Register
...........
.0. (set)
Empty Pipe
.8.
Urefn too large
MAX-Alarm
130 %
F115.67 %
F 6789.12 l