EasyManua.ls Logo

ABB COPA-XE - Interconnection Diagram MAG-XE

ABB COPA-XE
64 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
47
Electromagnetic Flowmeter
COPA-XE / MAG-XE
9.3.3 Interconnection Diagram MAG-XE
Connection Variations for Analog Communication (incl. HART)
1) a) Scaled Pulse Output, passive, pulse width can be set from 0.1 ms to 2000 ms,
Terminals: V8, V9, Function E9, C9,
Optocoupler specifications: f
max
5 kHz
0 V U
CEL
2 V, 16 V U
CEH
30 V
0 mA I
CEH
0.2 mA, 2 mA ≤ Ι
CEL
220 mA
b) Scaled Pulse Output, active pulse width can be set from 0.1 ms to 2000 ms,
Terminals V8, V9, Function 9, 10
20 mA < I 150 mA; f
max
4 Hz, pulse width50 ms, pulse T
16V
25 ms, 16 V U 30 V
On/off ratio 1:4 (T
on
: T
off
), f
max
5kHz, 2 mA I 20 mA; 16 V U 30 V
2) Contact Output, Function software selectable for System Monitoring, Empty Pipe,
Max.-Min.-Alarm or F/R Signal*, Terminals G2, P7
Optocoupler specifications:
0 V U
CEL
2 V, 16 V U
CEH
30 V;
0 mA I
CEH
0.2 mA, 2 mA I
CEL
220 mA
3) Contact Input, Function software selectable for External Zero Return or External Totalizer Reset,
Terminals: G2, X1
Optocoupler, 16 V U 30 V, Ri = 2k
4) Current Output selectable, Terminals: +/-, load 600 for 0/4 to 20 mA, load 1200 for 0/2 to 10 mA,
load 2400 for 0 to 5 mA
Option: HART-Protocol
5) Supply Power, see Instrument Tag
*) The Forward Direction Signal function is the factory setting.
Ground
PE
Converter
Flowmeter Primary
Ground
1) 2) 3) 4) 5)
6)
L < 50 m
Ground
Ground
Converter
Flowmeter Primary
Fig. 36 Interconnection Diagram MAG-XE for Analog Communication (incl. HART)

Table of Contents

Related product manuals