11
Electromagnetic Flowmeter
COPA-XE / MAG-XE
3.4. Parameter Overview and Data Entry in “Condensed Form”
Submenu/Parameter Entry Type Comments
from table/numeric Data can be entered only after the Program Protection has been
turned off.
on/off
If a number other than “0” ( factory setting) has been programmed
for the Program Protection Code, the Program Protection can only
be turned off after the correct PP-Code (1-255) has been entered.
Parameters can only be changed after the Program Protection
has been turned off.
numeric After the Program Protection is turned off it is also possible to
change the PP-Code.
Enter old PP-Code
0 = Factory setting
Enter new PP-Code (0-255)
from table German, English, French, Finnish, Spanish,
Italian, Dutch, Danish, Swedish.
For HART-Protocol only German, English, French,
Spanish
Present meter size
See Instrument Tag on flowmeter primary
Automatic setting based on the flowmeter size selection.
Flow range end value can be set from 0.05 to 1.0 Q
max DN
Flowmeter span value Cs for the selected
excitation frequency, see Instrument Tag
Flowmeter zero value Cz for the selected
excitation frequency, see Instrument Tag
ENTER
* Program Prot. *
off
PP-Code?
0
* Program Prot. *
off
ENTER
Old PP-Code?
0
New PP-Code?
0
ENTER
Meter Size
DN 250 10 In
Qmax DN 10 m/s
1800.00 m3/h
Span Cs 6.25 Hz
56.123 %
Zero Cz 6.25 Hz
0.1203 %