CoriolisMaster FCB400, FCH400 CORIOLIS MASS FLOWMETER | CI/FCB400/FCH400-EN REV. H 19
Inputs and outputs with optional plug-in cards Operating values (general) Type of protection – ‘nA’ / ‘NI’
U
N
I
N
U
N
I
N
Current output V3 / V4, active*
Terminals V3 / V4 and V1 / V2*
30 V 30 mA 30 V 30 mA
Current output V1 / V2, passive**
Current output V3 / V4, passive**
Terminals V1 / V2** or V3 / V4**
30 V 30 mA 30 V 30 mA
Digital output V3 / V4, active*
Terminals V3 / V4 and V1 / V2*
30 V 25 mA 30 V 25 mA
Digital output V1 / V2, passive**
Digital output V3 / V4, passive**
Terminals V1 / V2** or V3 / V4**
30 V 30 mA 30 V 30 mA
Digital input V3 / V4, active*
Terminals V3 / V4 and V1 / V2
30 V 3,45 mA 30 V 3,45 mA
Digital input V1 / V2, passive**
Digital input V3 / V4, passive**
Terminals V1 / V2** or V3 / V4**
30 V 3,45 mA 30 V 3,45 mA
Digital output V1 / V2, active*
Digital output V3 / V4, active*
Terminals V1 / V2** or V3 / V4**
24 V 22,5 mA 30 V 30 mA
Modbus® / PROFIBUS DP®
Terminals V1 / V2
30 V 30 mA 30 V 30 mA
Ethernet card
Ethernet (various protocols)
Port 1 / Port 2
57 V 417 mA 57 V 417 mA
Ethernet card in connection with Power over Ethernet (POE card)
Ethernet (various protocols)
Port 1 / Port 2
57 V 417 mA 57 V 417 mA
* Only in conjunction with additional ‘24 V DC loop power supply (blue)’ plug-in card in slot OC1.
** The terminal assignment depends on the model number or the slot assignments. For connection examples, see Connection examples on page 53.