EasyManua.ls Logo

ABB CoriolisMaster FCB400 Series - Housing as a Protective Device (Optional); Material Load for Process Connections; Material Load Curves for Flange Devices

ABB CoriolisMaster FCB400 Series
88 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
34 CoriolisMaster FCB400, FCH400 CORIOLIS MASS FLOWMETER | CI/FCB400/FCH400-EN REV. H
… 5 Installation
… Process conditions
Housing as a protective device (optional)
Order code PR5
Maximum burst pressure 60 bar (870 psi)
Optional order code PR6 and PR7 on request
Increased burst pressures up to 100 bar (1450 psi),
possible for nominal diameters DN 15 to 100 (½ to 4 in.).
Increased burst pressures up to 150 bar (2175 psi),
possible for nominal diameters DN 15 to 80 (½ to 3 in.).
Purge connections are available on request.
Pressure Equipment Directive
Conformity assessment in accordance with Category III, fluid
group 1, gas
Note the corrosion resistance of the meter tube materials in
relation to the measuring medium.
Material load for process connections
Note
You can reference the availability of the different process
connections in the Online ABB Product Selection Assistant (PSA)
for flow www.abb.com/flow-selector.
Not all connections shown here are available in all the devices
and designs.
The permissible material load of the device can additionally
differ from the material load of the connection. The
permissible limit values (pressure rating / measuring
medium temperature T
medium
) can be found on the name
plate.
Design Nominal diameter PS
ma
x
TS
ma
x
TS
min
Pipe fitting
(DIN 11851)
DN 15 to DN 40
(½ to 1½ in)
40 bar
(580 psi)
140 °C
(284 °F)
40 °C
(40 °F)
DN 50 to DN 100
(2 to 4 in)
25 bar
(363 psi)
140 °C
(284 °F)
40 °C
(40 °F)
Pipe fitting
(SMS 1145)
DN 25 to DN 80
(1 to 3 in)
6 bar
(87 psi)
140 °C
(284 °F)
40 °C
(40 °F)
Tri-Clamp
(DIN 32676)
DN 15 to DN 50
(½ to 2 in)
16 bar
(232 psi)
120 °C
(248 °F)
40 °C
(40 °F)
DN 65 to DN 100
(2½ to 4 in)
10 bar
(145 psi)
120 °C
(248 °F)
40 °C
(40 °F)
ASME BPE Clamp < DN 80
(< 3 in)
17.1 bar
(248 psi)
121 °C
(249.8 °F)
40 °C
(40 °F)
DN 80
(< 3 in)
15.5 bar
(224.8 psi)
121 °C
(249.8 °F)
40 °C
(40 °F)
DN 100
(< 4 in)
12.9 bar
(187.1 psi)
121 °C
(249.8 °F)
40 °C
(40 °F)
Material load curves for flange devices
Figure 12: Stainless steel DIN flange 1.4571 / 1.4404 (316Ti / 316L) to DN 200 (8 in)
Figure 13: Stainless steel ASME flange 1.4571 / 1.4404 (316Ti / 316L) up to DN 200 (8 in.)