CoriolisMaster FCB400, FCH400 CORIOLIS MASS FLOWMETER  |  CI/FCB400/FCH400-EN REV. H  49 
 
 
 
 
 
Current output Uco / 32 as loop power supply for digital 
output 41 / 42 or 51 / 52 
 
 
A Transmitter FCx400 
B Customer wiring 
OC1  Modbus / PROFIBUS DP plug-in
card 
R
B
 Load resistance 
Figure 31:  Current output Uco / 32 in power mode 
 
 
In the case of digital communication via Modbus / PROFIBUS DP, 
the current output Uco / 32 can be switched to the ‘Power Mode’ 
operating mode through the software. 
The current output 31/32/Uco is set permanently to 22.6 mA and 
no longer follows the selected process variable. HART 
communication is deactivated. 
As a result, the passive digital outputs 41 / 42 or 51 / 52 can also 
be operated as active digital outputs. 
 
The load resistance R
B
 needs to be integrated by the customer 
outside of the transmitter housing. 
 
Loop power supply 24 V DC pperating mode 
Terminals  Uco / 32
Function  For active connection of passive outputs
Output Voltage  Load dependent, see Figure 32.
Load rating I
max
  22.6 mA, permanently short circuit-proof
Table 1:  Specification current output Uco / 32 in power mode 
 
 
Figure 32:  Output voltage dependent on load resistance