52 CoriolisMaster FCB400, FCH400 CORIOLIS MASS FLOWMETER | CI/FCB400/FCH400-EN REV. H
… 6 Electrical connections
… Electrical data for inputs and outputs
Dependency of the output voltage U from the load R
B
.
Load R
B
is the parallel connection of the internal resistance R
int
and optional external resistance R
ext
.
Figure 39: Dependency of the output voltage U from the load R
B
.
Permissible ambient temperature single-compartment housing:
70 °C
Permissible ambient temperature dual-compartment housing
dependent on the load R
B
:
A Ambient temperature (°C)
Figure 40: Permissible ambient temperature dual-compartment housing
Binary output (active)
Terminals V1 / V2, V3 / V4
‘Off’ output U
L
200 mV
I
L
< 0 mA
‘On’ output 0 V U
H
24 V (dependent on R
B
)
0 mA I
H
22.5 mA (dependent on R
B
)
For pulse/frequency output
Terminals V3 / V4
‘Off’ output U
L
200 mV
I
L
< 0 mA
‘On’ output 0 V U
H
24 V (dependent on R
B
)
0 mA I
H
22.5 mA (dependent on R
B
)
f
max
10.5 kHz
Pulse width 0.05 to 2000 ms
Digital input V1 / V2, V3 / V4 (plug-in module)
A digital input can be implemented via the ‘Passive digital input
(yellow)’ plug-in module.
Can be configured as an input for external counter reset,
external output deactivation etc. via on-site software.
Figure 41: Plug-in card as digital input (I = internal, E = external)
The plug-in module can be used in slot OC1 or OC2.
Digital input
Terminals V1 / V2, V3 / V4
Input ‘On’ 16 V U
KL
30 V
Input ‘Off’ 0 V U
KL
3 V
Internal resistance R
i
6.5 kΩ
Function Can be configured using software.