EasyManua.ls Logo

ABB EL3060-Uras26 - Shewhart Control Charts for Quality Assurance; Shewhart Control Charts Functionality

ABB EL3060-Uras26
144 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EL3060 CONTINUOUS GAS ANALYZERS | OI/EL3060-EN REV. A 109
Shewhart control charts
Shewhart control charts are used to determine the combined precision and
drift of the AMS.
The differences d
t
between the measured values and the set points at the
zero point and the reference point are determined. These differences must
be less than or equal to the standard deviation s
AMS
multiplied by a factor of
2. If the difference for a measured value is greater, the measured value is
marked with the status "Out of Control".

Table of Contents

Other manuals for ABB EL3060-Uras26

Related product manuals