EL3060 CONTINUOUS GAS ANALYZERS  |  OI/EL3060-EN REV. A  137
 
 
 
Troubleshooting 
Flow error 
Gas lines or filter fouled, clogged up or leaky 
•  Disconnect gas analyzer from the gas conditioning system. 
•  Blow compressed air through gas lines or mechanically clear them. 
•  Exchange filter elements and filter beds. 
•  Check gas lines for seal integrity. 
Gas feed paths in the gas analyzer kinked or leaky 
•  Disconnect gas analyzer from the gas conditioning system. 
•  Check whether the gas lines in the gas analyzer are kinked or have be-
come detached from the connections. 
•  Check gas feed paths in the gas analyzer for seal integrity. 
Outlet pressure not the same as atmospheric pressure 
•  Ensure that the outlet pressure is the same as the atmospheric pressure. 
•  Conduct waste gases directly into the atmosphere or through a line with 
a large internal diameter which is as short as possible, or into a gas dis-
charge line. 
•  Do not conduct waste gases via restrictions or shut-off valves! 
Measured value display is not stable 
Vibrations 
•  Provide measures to reduce the vibrations. 
•  Increase the low pass time constant (see page 90). 
Leaks in the gas feed paths 
•  Check the gas feed paths for seal integrity (see page 117). 
Loss of sensitivity 
•  Check amplification drift display of the sample component (see page 121). 
Uras26: Emitter modulation uneven 
•  Have the emitter and modulator checked by service.