EL3060 CONTINUOUS GAS ANALYZERS | OI/EL3060-EN REV. A 121
Drift reading
Menu path
Maintenance Diagnosis Device Status Analyzer Status
Drift Indication
Drift
The offset and amplification drift values are calculated cumulatively starting
from the last basic calibration.
Delta drift
The offset and amplification delta drift values are calculated between the
last and last but one automatic calibration. They are deleted by a manual
zero-point or end-point calibration.
NOTES
The drift values are displayed in percentages of the customer measuring
range, i.e. the measuring range (see page 80) pre-set at the factory (see
analyzer data sheet (see page 33)).
The drift values are deleted by a calibration reset.