46 EL3060 CONTINUOUS GAS ANALYZERS | OI/EL3060-EN REV. A
Initial purging of gas feed paths
Initial purging of gas feed paths
Initial purging of the gas feed paths inside and outside the gas analyzer is
required before the power supply is switched on. A potentially explosive
gas/air mixture in the gas feed paths is thereby removed.
Purge gas for non-flammable sample gas Clean instrument air from
non-hazardous areas
Purge gas for flammable sample gas Nitrogen
Purge gas volume 5x volume of the gas feed paths
Purge gas flow approx. 30 l/h
Purging time at least 3 min
The purge gas for purging the EL3060-Uras26 analyzer unit may not contain
any fractions of the sample components.