EL3060 CONTINUOUS GAS ANALYZERS | OI/EL3060-EN REV. A 23
Sample gas inlet and outlet conditions for Magnos206, Magnos28, Caldos25,
Caldos27 analyzers
Temperature
+5 to +50 °C
Inlet and outlet pressure
The sample gas pressure in the sample gas feed path of the analyzer may be
max. 200 hPa positive pressure (1200 hPa absolute pressure). The pressure
drop at the flame barrier at the sample gas inlet means this can be achieved
by
Maintaining max. 200 hPa positive pressure (1200 hPa absolute pressure)
at the sample gas inlet or
Adhering to the pressure limits for the sample gas inlet and outlet as
shown in the following diagram:
Flow rate
Max. 80 l/h
Pressure drop at the flame barriers
Approx. 155 hPa at a flow rate of 50 l/h
Sample gas inlet and outlet conditions for Uras26 analyzer
Temperature
+5 to +45 °C
Inlet pressure
Absolute pressure max. 1200 hPa and gauge pressure max. 200 hPa
Flow rate
Max. 100 l/h
Pressure drop at the flame barriers
Approx. 40 hPa at a flow rate of 50 l/h