EL3060 CONTINUOUS GAS ANALYZERS | OI/EL3060-EN REV. A 69
Calibrate the sample component manually
Zero-point calibration
1 Select menu Zero point/Single point.
2 Select individual sample component or "All" (according to the configura-
tion in the ECT dialog "Manual Calibration").
3 Check zero-point set point
4
and adjust if required.
4 Connect zero reference gas (if it is not automatically connected).
5 Start calibration as soon as the measured value display is stable.
6 Store calibration or repeat calibration
5
.
End-point calibration
1 Select menu Span.
2 Select sample component.
3 Check span set point
6
and adjust if required.
4 Connect span gas (if it is not automatically connected).
5 Start calibration as soon as the measured value display is stable.
6 Store calibration or repeat calibration.
End-point calibration with calibration cell (option with Uras26)
1 Select menu Calibration Cell.
2 Select sample component or "All".
3 Connect zero reference gas (if not already connected).
4 Start calibration as soon as the measured value display is stable.
5 Store calibration or repeat calibration.
4
The parameterized test gas concentration is displayed. If the set point is altered
here, the parameterized test gas concentration is overwritten.
5
A calibration may have to be repeated if the measured value is not stable after initi-
ation of the calibration. The repeated calibration is based on the measured value
obtained in the preceding calibration.
6
The parameterized test gas concentration is displayed. If the set point is altered
here, the parameterized test gas concentration is overwritten.