D184B122U02 S4 25
Submenu/
Parameter
Submenu/
Parameter 
Submenu/
Parameter
Selections
Value Range
Comments
Prog Level
PP-Code
Old PP-Code: 1–9999
Normally the Prog. Protection Code is equal 
to "0". If the Prog. Prot. is selected, the 
question "New PP Code?" is displayed. If a 
code is entered between (1 and 9999) then 
an attempt to access the PP Code protect-
ed menus by an operator will result in the 
display of the message 
** PP Code ? ** indicating that the operator 
must enter the correct code before access 
to the menus is granted. To delete the 
PP Code it is necessary to access the 
submenu "PP Code". After pressing 
ENTER the "Old PP-Code?" is requested. 
After entering the correct value it is possible 
in the next submenu "New PP-Code?" to 
enter a new PP-Code value or by entering 
"0" to turn off the program protection.
ENTER
0
New PP-Code 1–9999
0
Submenu
Language 
Unit Qv
Unit Tv
Qmax          
Damping 1Tau       
Display mode 
D i s p l a y  1 s t  l i n e   
D i s p l a y  2 n d  l i n e      
Display 3rd line 
Display 4th line 
Iout type 
Iout Low Alarm     
Iout High Alarm        
Pulse factor
Pulse width 
With EASY -SETUP it is possible to config-
ure the metering system for operation using 
the fewest possible steps. The menus listed 
to the right are displayed for entry one after 
the other in "EASY-SETUP". If the selections 
1 big, 1 small are selected then the selection 
menus Display 3rd line, Display 4th line are 
not displayed. For HART instruments the 
selection “Iout type” is not displayed. Using 
ENTER the parameters can be changed one 
after the other, or if they are not to be 
changed, the C/CE key can be used to pro-
ceed to the next 
parameter. After making an entry with 
ENTER the entry can still be changed if 
necessary by using the ENTER Function 
and then using the STEP-key to proceed to 
the next parameter. For numeric entries, the 
minimum and maximum entry values are 
also displayed.
EASY SETUP
EASY SETUP
ENTER
Language 
German
English
French
Finish
Spanish
Italian
Dutch
Danish
Swedish
The languages available for the menu 
display are shown in the “Language” 
submenu.
Menus in the Service Code section use the 
English language.
English
Language
ENTER English
Submenu
Operating mode
Operating mode
Standard
Fast
Piston Pump
"Standard": selection used for continuous 
flow metering.
"Fast": accelerated measured value pro-
cessing (e.g. short batch operation"
“Piston Pump”: for pulsating flows. In this 
operating mode improved reproducibility of 
the measurements is achieved for pulsating 
flow in piston pump operation. The 
totalizer is automatically set for the operat-
ing mode Difference Totalizer. Any reverse 
flow which may occur is automatically de-
ducted from the totalizer value.
ENTER
Standard