Navigator 600 Series Analyzers
Silica and Phosphate Appendix B Acyclic Parameter Mapping
46 IM/NAV6PBS Issue 1
DIAGNOSIS_EXTENSION
1 0 Out of reagent 1 1
1 Out of reagent 2 1
2 Out of reagent 3 1
3 Out of reagent 4 1
4Out of all samples 1
5 Out of sample 1 1
6 Out of sample 2 1
7 Out of sample 3 1
2 0 Out of sample 4 1
1 Out of sample 5 1
2 Out of sample 6 1
3 Out of cleaning solution 1
4 Out of zero calibration solution 1
5 Out of secondary calibration solution 1
6Reserved for future use 0
7Reserved for future use 0
3 0 Reserved for future use 0
1Reserved for future use 0
2Reserved for future use 0
3Reserved for future use 0
4Reserved for future use 0
5Reserved for future use 0
6Reserved for future use 0
7Reserved for future use 0
4 0 Control temperature too high >10°C 1
1 Control temperature too high/low >2°C and <10°C 1
2 A/D error main measurement 1
3Reserved for future use 0
4 Faulty measurement optics 1
5 Failed reaction block temperature sensor 1
6 Excessive secondary current 1
7 Cleaning in progress 1
Octet Bit Description Diagnosis Mask (1 = enabled)
Table B.2 Diagnostic Data (Sheet 2 of 3)