56 VA Master FAM540 METAL CONE VARIABLE AREA FLOWMETER | OI/FAM540-EN REV. E
… 9 Operation
… Description of menus and parameters
‘Totalizer“ und „Overflow’ Menu
The number of counter overflows is indicated here. The
maximum number of counter overflows is 65535.
The overflow counter also overflows thereafter and resets to 0.
The total counter count can be calculated as follows:
Example:
Status of overflow counter: 12
Current counter status = 12345 m
3
12 x 10 000 000 = 120 000 / 000 m
3
+ 12 345 m
3
= 120 012 345 m
3
‘Damping’ Menu
The response time of the measuring device can be set as desired
using the Damping parameter.
Damping corresponds to a first-order low-pass filter.
Typical value 3 to 5 s.
10 Diagnosis / error messages
Calling up the error description
In accordance with NAMUR recommendation NE 107, a
distinction is made between error messages and warnings.
Errors and warnings are saved in the register.
Errors directly affect the current output.
Warnings can be read via the HART-protocol and processed.
Errors and warnings that have occurred can be called up by using
the ‘Status register’ Menu. When the menu is opened, the
number of errors and warnings is displayed. When the error
(warning) register is called up, the type of warning or error is
indicated.
1. From the Main Menu, select or to select the ‘Status
register’ Menu.
2. Confirm the selection with .
- Status re
g
ister -
Error register
No. 2
Bac
k
Edit
3. Open the first pending message with .
• Press to reset the message.
Error 3
Q > 103%
Cancel Reset
4. Open the additional pending messages with or .
• Press to reset the message.
Warning 1
Simulation
Cancel Reset
Note
Warning and error registers can be cleared on the ‘Standard’
user level to detect when they occur again.
The ‘Specialist’ user level has an additional ‘Mains interrupt’
submenu, where the number of power outages is recorded.
The power outage counter can only be reset by the ABB after-
sales service.