Table 11. Guidance and Manufacturer's Declaraon - Electromagnec Emissions
Emissions Test Compliance Electromagnec Environment
Guidance
Voltage uctuaons/ icker emissions
IEC 61000-3-3
power supply network that supplies
buildings used for domesc purposes.
Class 5 Full Compliance*
*Only applicable to Power Adapter (Model 16720)
Table 12. Guidance and Manufacturer's Declaraon - Electromagnec Immunity
Immunity Test IEC 60601
Test Level
Compliance Level Electromagnec
Environment Guidance
Electrostac
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
The relave humidity
should be at least 10%.
Electrical fast
transient/burst
IEC 61000-4-4*
±2 kV for power supply
lines
±1 kV for input/output lines
±2 kV for power supply
lines
±1 kV for input/output lines
Mains power quality should
be that of a typical
commercial, hospital, or
home environment.
Surge
IEC 61000-4-5*
±1 kV line to line
±2 kV line to ground
±1 kV line to line
±2 kV line to ground
Mains power quality should
be that of a typical
commercial, hospital, or
home environment.
Voltage dips, short
interrupons
IEC 61000-4-11*
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 s
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 s
Mains power quality should
be that of a typical
commercial, hospital, or
home environment. If the
user of the device requires
connued operaon during
power mains interrupons,
it is recommended that the
device be powered from an
uninterrupble power
supply or a baery.
Power frequency
(50/60 Hz) magnec eld
IEC 61000-4-8
30 A/m 30 A/m Power frequency magnec
elds should be at levels
characterisc of a typical
locaon in a typical
commercial, hospital, or
home environment.
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to
80 MHz
3 Vrms The device is intended for
use in the electromagnec
environment in which
radiated RF disturbances
are controlled. The
customer or the user of the
device can help prevent
electromagnec
interference by maintaining
a minimum distance
between portable and
mobile RF communicaons
equipment (transmiers)
and the device as
recommended below,
according to the maximum
output power of the
6 Vrms
ISM/Radio bands between
150 kHz to 80 MHz
80% AM at 1 kHz
6 Vrms
Radiated RF IEC 61000-4-3 10 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz
10 V/m
24