Table 11. Guidance and Manufacturer's Declaraon - Electromagnec Emissions
Emissions Test Compliance Electromagnec Environment
Guidance
Voltage uctuaons/ icker emissions
IEC 61000-3-3
power supply network that supplies
buildings used for domesc purposes.
Class 5 Full Compliance*
*Only applicable to Power Adapter (Model 16720)
Table 12. Guidance and Manufacturer's Declaraon - Electromagnec Immunity
Immunity Test IEC 60601
Test Level
Compliance Level Electromagnec
Environment Guidance
Electrostac
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
The relave humidity
should be at least 10%.
Electrical fast
transient/burst
IEC 61000-4-4*
±2 kV for power supply
lines
±1 kV for input/output lines
±2 kV for power supply
lines
±1 kV for input/output lines
Mains power quality should
be that of a typical
commercial, hospital, or
home environment.
Surge
IEC 61000-4-5*
±1 kV line to line
±2 kV line to ground
±1 kV line to line
±2 kV line to ground
Mains power quality should
be that of a typical
commercial, hospital, or
home environment.
Voltage dips, short
interrupons
IEC 61000-4-11*
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 s
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 s
Mains power quality should
be that of a typical
commercial, hospital, or
home environment. If the
user of the device requires
connued operaon during
power mains interrupons,
it is recommended that the
device be powered from an
uninterrupble power
supply or a baery.
Power frequency
(50/60 Hz) magnec eld
IEC 61000-4-8
30 A/m 30 A/m Power frequency magnec
elds should be at levels
characterisc of a typical
locaon in a typical
commercial, hospital, or
home environment.
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to
80 MHz
3 Vrms The device is intended for
use in the electromagnec
environment in which
radiated RF disturbances
are controlled. The
customer or the user of the
device can help prevent
electromagnec
interference by maintaining
a minimum distance
between portable and
mobile RF communicaons
equipment (transmiers)
and the device as
recommended below,
according to the maximum
output power of the
6 Vrms
ISM/Radio bands between
150 kHz to 80 MHz
80% AM at 1 kHz
6 Vrms
Radiated RF IEC 61000-4-3 10 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz
10 V/m