Table 12. Guidance and Manufacturer's Declaraon - Electromagnec Immunity
Immunity Test IEC 60601
Test Level
Compliance Level Electromagnec
Environment Guidance
communicaons
equipment.
Recommended minimum
separaon distance for
higher immunity test levels
where P is the maximum
power in was (W), d is the
recommended separaon
distance in meters (m), and
E is the immunity test level
in volts per meter (V/m).
*Only applicable to Power Adapter (Model 16720)
NOTE:
U
T
is the AC mains voltage prior to applicaon of the test level.
These guidelines may not apply in all situaons. Electromagnec propagaon is aected by absorpon and
reecon from structures, objects, and people.
Table 13. Guidance and Manufacturer's Declaraon - Proximity Fields from RF
Wireless Communicaon Equipment
Proximity Test IEC 60601 Test Level Compliance Level Electromagnec
Environment Guidance
IEC 61000-4-3 385 MHz: 27 V/m @ 18 Hz
pulse modulaon
27 V/m Recommended minimum
separaon distance
d = 0.3 m
450 MHz: 28 V/m @ FM
modulaon
28 V/m
710 MHz, 745 MHz,
780 MHz: 9 V/m @ 217 Hz
pulse modulaon
9 V/m
810 MHz, 870 MHz,
930 MHz: 28 V/m @ 18 Hz
pulse modulaon
28 V/m
1720 MHz, 1845 MHz,
1970 MHz: 28 V/m @
217 Hz pulse modulaon
28 V/m
2450 MHz: 28 V/m @
217 Hz pulse modulaon
28 V/m
5240 MHz, 5500 MHz,
5785 MHz: 9 V/m @ 217 Hz
pulse modulaon
9 V/m
WARNING: Do not use portable RF communicaons equipment (including peripherals such as antenna cables
and external antennas) closer than 30 cm (12 inches) to any part of the device, including cables specied by
Abbo Medical. Otherwise, performance degradaon may occur.
25