Table 12. Guidance and Manufacturer's Declaraon - Electromagnec Immunity
Immunity Test IEC 60601
Test Level
Compliance Level Electromagnec
Environment Guidance
communicaons
equipment.
Recommended minimum
separaon distance for
higher immunity test levels
where P is the maximum
power in was (W), d is the
recommended separaon
distance in meters (m), and
E is the immunity test level
in volts per meter (V/m).
*Only applicable to Power Adapter (Model 16720)
NOTE:
• U
T
is the AC mains voltage prior to applicaon of the test level.•
•
These guidelines may not apply in all situaons. Electromagnec propagaon is aected by absorpon and
•
reecon from structures, objects, and people.
Table 13. Guidance and Manufacturer's Declaraon - Proximity Fields from RF
Wireless Communicaon Equipment
Proximity Test IEC 60601 Test Level Compliance Level Electromagnec
Environment Guidance
IEC 61000-4-3 385 MHz: 27 V/m @ 18 Hz
pulse modulaon
27 V/m Recommended minimum
separaon distance
d = 0.3 m
450 MHz: 28 V/m @ FM
modulaon
28 V/m
710 MHz, 745 MHz,
780 MHz: 9 V/m @ 217 Hz
pulse modulaon
9 V/m
810 MHz, 870 MHz,
930 MHz: 28 V/m @ 18 Hz
pulse modulaon
28 V/m
1720 MHz, 1845 MHz,
1970 MHz: 28 V/m @
217 Hz pulse modulaon
28 V/m
2450 MHz: 28 V/m @
217 Hz pulse modulaon
28 V/m
5240 MHz, 5500 MHz,
5785 MHz: 9 V/m @ 217 Hz
pulse modulaon
9 V/m
WARNING: Do not use portable RF communicaons equipment (including peripherals such as antenna cables
and external antennas) closer than 30 cm (12 inches) to any part of the device, including cables specied by
Abbo Medical. Otherwise, performance degradaon may occur.
25