Table 14. Guidance and Manufacturer's Declaraon - Enclosure Port Immunity to RF
Wireless Communicaon Equipment
Immunity Test Band (MHz) Maximum Power (W) Immunity Test Level (V/m)
IEC 61000-4-3 380 to 390 1.8 27
430 to 470 2 28
704 to 787 0.2 9
800 to 960 2 28
1700 to 1900 2 28
2400 to 2570 2 28
5100 to 5800 0.2 9
Table 15. Guidance and Manufacturer's Declaraon - Enclosure Port Immunity to
Proximity Magnec Fields
Immunity Test Test Frequency Modulaon Immunity Test Level (A/m)
IEC 61000-4-39 30.0 kHz Connuous wave 8
134.2 kHz Pulse modulaon 2.1 kHz 65
13.56 MHz Pulse modulaon 50 kHz 7.5
Appendix E: Symbols and Denions
The symbols below and harmonized symbols may be found on the product or product
label. For harmonized symbols, refer to the Universal Symbols Glossary at
medical.abbo/manuals.
Table 16. Symbols and denions
Symbol Denion
Cauon
Refer to instrucon manual/booklet
Consult instrucons for use
Follow instrucons for use on this website
Type BF Applied Part
MR Unsafe