wear period. Customer Service is available at 1-855-632-8658 7 Days a Week
from 8AM to 8PM Eastern Standard Time.
Do not reuse Sensors. The Sensor and Sensor Applicator are designed for single
use. Reuse may result in no glucose readings and infection. Not suitable for re-
sterilization. Further exposure to irradiation may cause unreliable low results.
If a Sensor breaks inside your body, call your health care professional.
How to Store the Sensor Kit:
Store the Sensor Kit between 36°F and 82°F. Storage outside of this range may
cause inaccurate Sensor glucose readings.
If you suspect that the temperature may exceed 82°F (for example, in an un-
airconditioned home in summer), you should refrigerate your Sensor Kit. Do not
freeze your Sensor Kit.
Store your Sensor Kit in a cool, dry place. Do not store your Sensor Kit in a
parked car on a hot day.
Store the Sensor Kit between 10-90% non-condensing humidity.
When not to use the System:
Do NOT use if the Sensor Kit package or Sensor Applicator appear to be
damaged or if tamper label indicates Sensor Applicator has already been
opened.
Do NOT use if Sensor Kit contents are past expiration date.
What to know about the System:
The FreeStyle Libre 3 System is intended for use by a single person. It must not
be used by more than one person due to the risk of misinterpreting glucose
information.
What to know before you Apply the Sensor:
Wash application site on the back of your upper arm using a plain soap, dry, and
then clean with an alcohol wipe. This will help remove any oily residue that may
prevent the Sensor from sticking properly. Allow site to air dry before
DOC44383_rev-C 1:14pm 22-SEP-2021

Table of Contents

Other manuals for Abbott FreeStyle Libre 3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott FreeStyle Libre 3 and is the answer not in the manual?

Abbott FreeStyle Libre 3 Specifications

General IconGeneral
Sensor TypeContinuous Glucose Monitoring (CGM)
Sensor Wear Duration14 days
ConnectivityBluetooth
App CompatibilityFreeStyle Libre 3 app (iOS and Android)
Data Update FrequencyEvery minute
Warm-up Time60 minutes
AccuracyMARD (Mean Absolute Relative Difference) of 7.9%
CalibrationFactory calibrated
Battery LifeSensor lasts for 14 days, no battery replacement required
Data Storage8 hours of data stored on the sensor
AlertsHigh and low glucose alerts
Measurement Range40 to 400 mg/dL (2.2 to 22.2 mmol/L)