EasyManua.ls Logo

Abbott FreeStyle Libre 3 - Accuracy to YSI Within CGM Glucose Ranges

Abbott FreeStyle Libre 3
97 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
* No YSI measurements were obtained for children ages 4-5; results displayed are from CGM-SMBG matched paired measurements.
The accuracy of different CGM glucose ranges versus YSI reference was assessed
by calculating the percentage of System readings that were within 15%, 20%, and
40% for reference values 70 mg/dL, and within 15 mg/dL, 20 mg/dL, and
40 mg/dL for values < 70 mg/dL. For blood glucose values < 70 mg/dL, the
difference in mg/dL between the CGM and YSI blood glucose values was
calculated. For values 70 mg/dL, the relative difference (%) to the YSI blood
glucose values was calculated. The results categorized within CGM glucose ranges
are presented in Tables 1b and 1c. The results categorized within YSI glucose
ranges are presented in Tables 1d and 1e.
Table 1b: Accuracy to YSI within CGM Glucose Ranges (Adult; n=200)
System range is 40-400 mg/dL.
Table 1c: Accuracy to YSI within CGM Glucose Ranges (Pediatric*; n=168)
DOC44383_rev-C 1:14pm 22-SEP-2021