* Includes children 6-17 years of age. No YSI measurements were obtained for children 4-5 years of age.
System range is 40-400 mg/dL.
Table 1d: Accuracy to YSI within YSI Glucose Ranges (Adult; n=200)
Table 1e: Accuracy to YSI within YSI Glucose Ranges (Pediatric*; n=168)
* Includes children 6-17 years of age. No YSI measurements were obtained for children 4-5 years of age.
Agreement with ‘LO’ and ‘HI’ CGM Reading against YSI Reference
The System reports glucose concentrations between 40 and 400 mg/dL. When the
System determines that glucose level is below 40 mg/dL, it will report as ‘LO’.
When the System determines that glucose level is above 400 mg/dL, it will report
as ‘HI’. Tables 2a and 2b display the concurrence between the CGM and YSI
DOC44383_rev-C 1:14pm 22-SEP-2021