Abbott Freestyle Fredom Specifications

General IconGeneral
ModelFreeStyle Freedom
Sample Size0.3 µL
DisplayLCD
Measurement Range20 to 500 mg/dL
Battery1 x CR2032
Operating Temperature41°F to 104°F (5°C to 40°C)
Relative Humidity10% to 90%

Summary

Important Information About Your FreeStyle Freedom™ Blood Glucose Monitoring System

How the FreeStyle System Should Be Used

Explains the intended use, sample sources, and compatible accessories for the system.

Important Health-Related Information

Health and Safety Precautions

Covers health conditions, side effects, and factors influencing blood glucose test results for safe usage.

The FreeStyle Freedom™ Meter and How It Works

Meter Components Overview

Describes key parts of the FreeStyle Freedom meter, including data port, display, and buttons.

Test Strip Application Area

Details where to apply blood or control solution on the test strip edge.

The FreeStyle Freedom™ Meter Display

System Check and Status Indicators

Explains the initial system check screen and various status symbols like memory, battery, and sound.

The FreeStyle Freedom™ Meter Display

Measurement and Result Display Areas

Details the areas on the meter display for test results, units of measurement, and averages.

The FreeStyle Freedom™ Meter Display

Symbol Meanings for Meter Operation

Clarifies symbols related to test results, codes, averages, and messages displayed on the meter.

The FreeStyle® Test Strip

Test Strip Usage and Handling

Provides instructions on applying samples, warnings, and essential information for test strip care and storage.

The FreeStyle® Control Solution

Control Solution Purpose and Information

Explains the function of control solution for practice and system verification, plus usage guidelines.

Performing a Control Solution Test

Step-by-Step Control Test Procedure

Guides users through the process of conducting a control solution test and interpreting results.

The FreeStyle® Lancing Device

Using the Lancing Device for Blood Samples

Details how to set up, load, and operate the lancing device for obtaining blood samples.

Testing Your Blood Glucose

Blood Glucose Testing Preparation and Timing

Covers essential preparation like hand washing, site selection, sample size, and recommended testing times.

Performing a Blood Glucose Test

Meter Preparation and Test Strip Coding

Outlines the initial steps of preparing the meter and setting the correct test strip code for accurate readings.

Performing a Blood Glucose Test

Lancing Test Sites and Obtaining Samples

Explains how to lance different body sites and collect the correct blood sample size.

Performing a Blood Glucose Test

Filling Test Strip and Waiting for Results

Guides on correctly filling the test strip with blood and the process of waiting for the glucose reading.

Performing a Blood Glucose Test

Lancet Removal and Meter Maintenance

Details how to safely remove the lancet after use and basic meter cleaning instructions.

LO and HI Readings

Interpreting Low and High Glucose Readings

Explains the meaning of LO and HI results, their implications, and recommended actions.

Meter Set-Up

Initial Meter Configuration

Covers entering setup mode and navigating through initial settings like alarms.

Setting Alarms and Notifications

Provides instructions on how to set up to four different alarms for blood glucose monitoring.

Meter Set-Up

Configuring Time and Alarm Volume

Details how to set the meter's hour, minute, and alarm volume levels.

Meter Set-Up

Setting Time, Date, and Format Options

Guides users on setting the time, date, and choosing between 12/24-hour and various date formats.

Meter Set-Up

Date Format and Unit of Measurement

Explains how to set the date format and identifies the unit of measurement used by the meter.

Meter Set-Up

Software Version and Setup Completion

Displays the meter's software version and confirms the completion of the meter setup process.

Meter Results

Accessing Meter Memory and Averages

Explains how to enter meter memory to view the 14-day average and stored test results.

Understanding Test Result Displays

Details how blood glucose, control solution, and temperature-affected results are shown on the meter.

Meter Results

Interpreting Specific Test Result Types

Differentiates between blood glucose tests, control solution tests, and explains the "PC" symbol for data transfer.

Meter Maintenance

Battery Information and Replacement

Provides details on the meter's battery, its life, and the procedure for replacing it.

Meter Maintenance

Lancing Device and Meter Care

Offers guidance on cleaning the lancing device and the exterior of the meter to ensure proper function.

System Troubleshooting

Troubleshooting Error Messages

Lists common error codes (Error 1-6), their possible causes, and recommended actions to resolve them.

System Troubleshooting

Addressing Error 1 and Error 2

Details the causes and solutions for Error 1 (small sample, strip/meter issue, low glucose) and Error 2 (high glucose, cold temp control solution).

System Troubleshooting

Resolving Error 3 and Error 4

Explains causes and solutions for Error 3 (incorrect procedure, strip/meter issue) and Error 4 (strip/meter issue).

System Troubleshooting

Handling Error 6 and General Issues

Covers Error 6 (button press during sample application) and notes Error 5 is unused, plus general troubleshooting steps.

System Troubleshooting

Troubleshooting Common Meter Issues

Provides a table addressing issues like meter not powering on or test not starting, with causes and solutions.

FreeStyle Freedom™ System Specifications

Technical Specifications Overview

Lists key technical details of the system, including assay method, battery life, range, and dimensions.