39
ENGLISH
FreeStyle Freedom™ System Specifications
Assay Method Coulometric electrochemical sensor
Automatic Shutoff Two minutes after last user action
Battery Life 1,000 tests
Calibration Plasma equivalent
Hematocrit 15% to 65%
Measurement Units mg/dL (
milligrams per deciliter)
Meter Storage Temp. -4º to 140º F (-20º to 60º C)
Memory 250 blood glucose and contro
l solution
tests with date and time
Operating Relative 5% to 90% (non-condensing)
Humidity
Operating Temperature
40º to 104º F (5º to 40º C)
Power Source One #2032 battery, 3 volt lithium replaceable
Result Range 20 to 500 mg/dL
Sample Whole blood, capillary
Sample Size 0.3 microliter (300 nanoliters)
Size Width 2.00” x Leng
th 3.30” x
Thickness 0.63” (at thickest point)
Test Time Average
of 5 seconds
Weight 1.43 oz. (including battery)
Probable Cause What to Do
Blood sample is too small. You may add blood to the Test Strip Edge for up to 60 seconds
from the time of the first application. If necessary, perform
another test using a new test strip and a larger blood sample.
Defective test strip. Repeat the test with a new test strip.
Sample applied after automatic shutoff Repeat the test using a new test strip. Wait until you see the
(two minutes). Blood Drop and Test Strip Symbols on the display screen before
you apply the blood sample
Defective meter or test strips. Call Customer Care.
ART06690-Rev-B-EN.indd 39 11/17/05 6:12:06 PM