51
6. Squeeze the bottle until a tiny drop forms
at the tip. Touch the drop to the front
edge of the yellow window of the test
strip. When you see flash, you have
enough control solution in the test strip.
Wipe the tip of the bottle with a tissue,
then cap the bottle tightly.
7. The result appears on the display, along
with the control bottle symbol and flashing
“L.” Do not remove the test strip yet.
Press once to mark the result as a
Level 1. If you tested the Level 2 control,
press a second time.

Other manuals for Accu-Chek Aviva

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Aviva and is the answer not in the manual?

Accu-Chek Aviva Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
CodingNo coding required
Battery LifeApproximately 1000 tests
Memory500 test results
Battery Type1 x 3V lithium (CR2032)
Operating Humidity10% to 90% RH
Meter Weight59 g
Measuring Range10 mg/dL to 600 mg/dL
Automatic Shut OffAfter 90 seconds of inactivity