53
Range (mmol/L)
Level 1 1.4–3.1
Level 2 14.2–19.1
Understanding Control Test Results
The label on the test strip container shows
the acceptable ranges for both the Level 1
and Level 2 control solutions. The result you
get should be inside this range. Make sure
you compare the result to the correct level
of control.
When the control result is inside the range
on the test strip container, the test strips
and meter are working properly.
Example