When the self-test is finished, the INFUSION SET
PRIMING screen is displayed.
Press f
and move to 3.11 priming the infusion
set step 2.
11.
10.
Press f after correctly inserting the cartridge.
Your Pump performs a self-test.
Check that the infusion set luer-lock connector is
still correctly connected to the adapter. Tighten
the infusion set by hand clockwise until it sits
tightly in the adapter.
9.
Incorrect use can cause damage to your insulin pump
Do NOT hold your Pump upright when inserting the
cartridge. This could lead to insulin flowing into the
cartridge compartment and cause a blockage of the
piston rod. The Pump may then display an E6 or E10
error message or stop.
Do NOT insert the cartridge into the insulin pump
without the infusion set tubing connected. This could
lead to insulin flowing into the cartridge compartment.
If insulin gets into the compartment this can result in a
damage of the piston rod over time and potentially limit
or cause a blockage of the insulin pump motor functions.