Typical Range and Maximum Parameter
Parameter Settings Range
Basal Rate
hourly basal rate + U100: 62.5 IU/h Insulin Type Max. Amount
max. TBR U100 0 – 62.50 IU/h
(hourly basal rate
combined with temporary
basal rate, max. increase)
hourly basal rates 0.1 U
increment
hourly basal rates U100: 10.0 IU/h Insulin Type Max. Amount
U100 0 25.0 IU/h
temporary basal rate 100% 0 – 200%
TBR increase or decrease
10% 10%
in increments of
TBR duration max. 24 h 15 min – 24 h
TBR duration increments 15 min 15 min, 30 min, 1 h
basal rate profiles available 5 1 – 5
basal ra
te lock off on or off
Bolus
bolus amount per U100: 25.0 IU/h Insulin Type Max. Amount
bolus delivery U100 0 – 25.0 U
bolus amount increments 0.5U 0.1, 0.2, 0.5, 1.0 or 2.0 U
for “Quick” and “Scroll”
Standard Bolus
bolus duration (Extended max. 24 h 15 min – 24 h
bolus, MultiWave bolus)
bolus duration increments 15 min 15 min, 30 min, 1h
Appendixes ACCU-CHEK Spirit Insulin Pump178