List of Errors
Error E1: CARTRIDGE EMPTY
The insulin cartridge is empty. Change the cartridge immediately.
1. Press f twice to turn off and to confirm the error.
2. Disconnect or remove the infusion set from your infusion site.
3. Change the cartridge and prime the infusion set.
4. Reconnect the infusion set at your infusion site.
5. Put your pump in RUN mode.
See the section “Changing the Cartridge and Infusion Set” (page 73) for more
information.
!
After the occurrence of an error, your insulin pump will be in
STOP mode and the insulin delivery is interrupted. In order
to maintain insulin delivery, you must act immediately
according to the instructions given for each error code.
Chapter 7: Alerts and Errors ACCU-CHEK Spirit Insulin Pump154
WARNING