Follow the steps below:
1.
From the RUN screen press and hold s (a) until you hear a beep
sequence and feel a vibration.
This activates the “Quick” STANDARD
BOLUS function.
2. Press a (s) repeatedly until the bolus amount is reached.
Each press of the
a (s) key adds one bolus
increment to the bolus amount. Your pump
beeps and vibrates simultaneously once for
each programmed bolus increment.
Five seconds after the last press of
a (s), your pump confirms the total
bolus amount with one beep and vibration for each bolus increment programmed.
The Standard Bolus symbol ( ) blinks for 5 seconds (the bolus delivery
start delay).
After this bolus deliver
y start delay, the pump beeps three times and then
begins to deliver the total bolus amount programmed. The countdown of the
remaining bolus amount appears on the display.
Hint:
Start programming with the a key in areas of poor lighting, as this
will also turn on the backlight.
RUN
screen
bolus
amount
confirmation
of amount
3 sec.
5 sec. delay
3 sec.
if O unit
bolus
delivery
A8: bolus cancelled
5 sec. delay