Make sure that the cancellation was intended and then program a new
Extended Bolus (and/or Temporary Basal Rate), if needed.
Note:
If a Temporary Basal Rate is active at the same time, it is also cancelled
when your pump is put into STOP mode. An alert A8: BOLUS CANCELLED
and an alert A6: TBR CANCELLED occur.
Press
f twice to confirm and turn off the first alert. The second alert
appears on the display. Press f twice to confirm and turn off the second
alert. Both alerts are recorded in the alarm history.
See the sections “Alert A6: TBR CANCELLED” (page 152) and “Alert A8:
BOLUS CANCELLED” (page 153) for more information.