Ultrasonic flaw detector А1212 MASTER
Operation Manual
36
ACOUSTIC
CONTROL
SYSTEMS
2
2.3.1.6 System instrument settings
The system settings are common for all operating modes of the instrument.
The screen used for the system settings is presented in the Figure 32.
The names of the parameters and their permissible values are presented in the Table 16.
Figure 32
Parameter name Value Description
Firmware version X.X.X Current firmware version
Time HH:MM
Current time in HOURS:MINUTES 24 hour format.
Enters the editing mode -
Date DD.MM.YYYY
Current date in DAY.MONTH.YEAR format.
Enters the editing mode -
Brightness
from 5 to 100 with
step 5
Setting image brightness of the instrument’s
screen
Sound On/Off
Switch On/Off the accompanying sound of AFAS
actuation by the instrument and connected
Bluetooth headset
Auto-shutdown
timer, min
5 / 10 / 15 / 30 / 60
Selecting a time interval at the end of which
the instrument will switch off automatically if
neither key is pressed nor measurements are
performed within this period.
Free memory, Mb ХХХХ Free memory volume
Table 16