21
•Possible interference.
Check to see if the central unit has been positioned too close to another electrical
device which can cause interference. In this case, move the interfering electrical
device further away from the central unit.
•Broken connections inside.
•Display damaged.
•PCB damaged.
If you cannot find the cause of the malfunction, immediately discontinue use of
the device and contact SATELEC after-sales service.
Fault detected: The keyboard is not working properly
Possible cause
Solution
•Keyboard damaged.
•Broken connections inside.
Attempt to change parameters by using the footswitch.
Contact SATELEC after-sales service.
Fault detected: The sound signals are not working properly
Possible cause
Solution
•Broken connections inside
Immediately discontinue use of the device and contact SATELEC after-sales
service.
Fault detected: The motor doesn’t work
Possible cause
Solution
• Micro motor cable not connected properly
Check to see that the micro motor cable is plugged into the connector on the
front right of the central unit, disconnect and reconnect.
•Contra-angle defective or not properly
calibrated.
Check the integrity of the Contra-angle. Remove the Contra-angle and perform
calibration.
• Footswitch damaged.
Attempt to start the Micro Motor by using the
[Motor ON/OFF button]
on the
central unit keyboard. If unit does not turn on, contact SATELEC after-sales
service.
•Wrong voltage.
Verify that the voltage matches the power supply. If it does, try to switch the
central unit off and then on again, or connect the device to another electrical
outlet.
•PCB damaged.
•Motor damaged.
•Connector pins and/or cable damaged.
Immediately discontinue use of the device and contact SATELEC after-sales
service.
Fault detected: The motor loses power.
Possible cause
Solution
•Contra-angle defective or not properly
calibrated.
Check the integrity of the Contra-angle by removing the Contra-angle and perform
calibration.
•Motor damaged.
Contact SATELEC after-sales service.
Fault detected: The motor stops suddenly
Possible cause
Solution
•Micro motor/Cable damaged
Check to see that the micro motor cable is plugged into the connector on the
front right of the central unit, disconnect and reconnect. Start the motor by
pressing the
[Motor ON/OF] button on the
keyboard. If the motor turn on,
immediately discontinue use of the device and contact SATELEC after-sales
service.
•Wrong voltage.
Verify that the voltage matches the power supply. If it does, try to switch the
central unit off and then on again, or connect the device to another electrical
outlet.
• Footswitch damaged.
Attempt to start the Micro Motor by using the
[Motor ON/OFF button]
on the
central unit keyboard.
If unit does not turn on, and contact SATELEC after-sales service.