•PCB damaged.
If you cannot find the cause of the malfunction, immediately discontinue use of
the device and contact dealer, the authorized service center, or the manufacturer
directly.
Fault detected: The Motor will not turn off
Possible cause
Solution
•Keyboard and/or foot pedal damaged
Try to turn off the motor by pressing the
[Motor ON/OFF] button
on the keyboard
or by using the foot pedal.
If the motor will not turn off, immediately discontinue use of the device and
contact SATELEC after-sales service.
•Broken connections inside.
•PCB damaged
Immediately discontinue use of the device and contact SATELEC after-sales
service.
•Wrong voltage.
Verify that the voltage matches the power supply. If it does, try to switch the
central unit off and then on again, or connect the device to another electrical
outlet.
Fault detected: The pump is not working property
Possible cause
Solution
•Rotor is blocked.
Disconnect power to the central unit. Check to see if the rotor of the peristaltic
pump rotates correctly and is not blocked by any internal or external objects.
Remove any visible blockage if possible. Otherwise, contact SATELEC after-sales
service.
•Tube not correctly inserted
•Tube damaged.
To verify the proper tube connection follow the instructions on Chapter VIII
USING THE DEVICE.
Check for damage along the tube, kinks, partial occlusions, etc. Arrange for
immediate replacement of the tube.
Fault detected: The pump stops suddenly
Possible cause
Solution
•Rotor is blocked.
Disconnect power to the central unit. Check to see if the rotor of the peristaltic
pump rotates correctly and is not blocked by any internal or external objects.
Remove any visible blockage if possible. Otherwise, contact SATELEC after-sales
service.
•Pump damaged/blocked.
•PCB damaged.
Immediately discontinue use of the device and contact SATELEC after-sales
service.
XII ELECTROMAGNETIC COMPATIBILITY
Important: The power cord, micro-motor cord, and footswitch cord must be kept apart.
The I-SURGE requires special precautions to be taken with regard to electromagnetic compatibility. It must be
installed and prepared for use as described in chapter IV.
Certain types of mobile telecommunication devices such as mobile telephones are likely to interfere with the I-
SURGE. The recommended separation distances in this paragraph must therefore be complied with.
The I-SURGE must not be used near to or on top of another device. If this cannot be avoided, its operation under
the conditions of use must be checked beforehand.