18
Error Displays and Troubleshooting
ERROR SYMBOL/SYMPTOM CAUSE REMEDY
OVER-PRESSURIZATION:
The cuff was inflated to the
maximum pressure because
of movement of body, etc.
MEASUREMENT ERROR:
Measurement could not be
made because of moving or
talking during measurement.
INFLATION ERROR: AIR PLUG
is not correctly inserted.
Noise was detected.
DEFLATION ERROR:
Movement or talking was de-
tected during measurement.
Extreme changes in pressure
occurred during measurement.
CHECKING MODE:
See page 21.
NO POWER:
AC ADAPTER is not
correctly connected.
Battery is exhausted.
9002 was left turned on but
has not been operated for over
an hour.
Make sure that the patient stays
still during measurement.
Make sure the patient remains
still and quiet during measure-
ment.
Reinsert AIR PLUG and make
sure that it is securely inserted.
Make sure that the patient stays
still and quiet during
measurement.
Make sure the patient remains
still and quiet during
measurement.
Do not let patient move during
measurement.
Turn off the device once and
then turn it on again to take
measurement.
Reinsert AC ADAPTER and
make sure that it is securely
inserted.
Charge the battery or use AC
ADAPTER.
Turn off the device or press
START/STOP button or
SET/MEMORY button to acti-
vate the device from energy
saver mode.
Nothing is being dis-
played when MAIN-
POWER SWITCH is
pressed down in the ON
position.