21
Because the device includes precision parts, care should be taken to avoid ex-
treme temperature variations, humidity, shock, dust, and direct sunlight. Do not
drop or strike the device. Make sure not to expose the unit to moisture. This
unit is not water resistant.
Use only a soft, dry cloth to clean the device. Do not use gasoline, paint thinner,
chemicals such as strong bases, strong acids, oxidizing agents, and reducing
agents, or other strong solvents on the unit. Since the cuffs may absorb perspi-
ration and other fluids, inspect them for stains and discoloration after each use.
Sponge with a damp cloth. If necessary the cuff can be washed in cool water
with mild soap or detergent-disinfectant. Remember to remove the bladder prior
to washing. After washing, the cuff should be rinsed and allowed to air dry.
NEVER iron.
When storing the device, do not place heavy objects on it and do not coil BLADDER
TUBE too tightly. When the unit has been stored at a temperature below the
freezing point, keep it for at least 1 hour in a warm place before using. Remove
the battery if the device is to be stored for an extended period of time. Keep the
batteries out of the reach of children.
Caution: Do not inflate the cuff when it is not wrapped around an arm.
Caution: Do not disassemble or modify the device.
We suggest that the device be checked every 2 years. This operation may only
be performed by the manufacturer or by firms authorized by the manufacturer.
Checking pressure accuracy
Connect a cuff wrapped around a firm cylinder to the 9002
e-sphyg
TM
2 and a standard manometer. Turn on the 9002
e-sphyg
TM
2 while pressing and holding the START/STOP but-
ton down. Release the button after the initial display (all segment
display) disappears. Two "0"s are displayed and the 9002
e-
sphyg
TM
2 is in CHECKING MODE. Inflate the cuff with an infla-
tion bulb and compare the readings. Pressure accuracy is to be
±3mmHg. Contact dealer if accuracy of your 9002
e-sphyg
TM
2
is invalid.
(A test kit may be purchased from an authorized dealer)
9002
e-sphyg
TM
2
9002
PORT
BLADDER
TUBE
Care and Maintenance