17
1. In main display, long press Unit key to switch the display items in the sub display area, and switch to the Tare
setting item, as shown in Figure 8. The Tare icon does not flash under the Tare setting item, and the sub display shows
OFF to indicate that the Tare function is turned off.
Figure 8 Tare settings
2. At this moment, short press the Zero button, and the gauge will capture the current pressure value as the Tare
value, which will be displayed in the sub display area, and the current pressure value will be zeroed, as shown in Figure 9.
Short press the zerobutton again to turn off the Tare function. When the Tare is enabled, long press the Unit button to
switch to other setting items in the sub display area, then Tare icon will flash, indicating that Tare is enabled.