30
Available for first-order filtering mode, default is 0.05, can
be changed by commands
Number of filtered
samples
Number of samples for average filtering
Extreme value pairs will
be removed
0
~
4 (The Extreme value pairs number
should not exceed (Samples number-1)/2)
Extreme value pairs will be removed in average filtering,
default is 0, can be changed by commands
3.1.7 Temperature Unit Settings
The title bar displays 1-7, the main display area displays TEMP, short press the Enter button to enter the temperature
unit settings. It supports Celsius and Fahrenheit unit switching.
3.1.8 Pressure Type
The title bar displays 1-8, the main display area displays P-TYPE, short press the Enter button to switch between
gauge pressure (GAGE) and absolute pressure (AbS).
3.2 Custom UnitSettings
The title bar displays 2-1, the main display area displays COE, short press the Enter button to enter the custom unit
settings. User-defined unit coefficients can be modified by commands, and can also support selectinH2O(20
℃
),inH2O(60
℉
),mmH2O(20
℃
),mmH2O(15
℃
),ftH2O(60
℉
),ftH2O(4
℃
) as Custom units, as shown below: