20
To Clean CSS1 Optical Port
If using AFL CCTS cleaning tips and FCC2 cleaning uid:
1 Leaning the FCC2 can back (30°), press the button on the FCC2 to ll the well.
2 Dip the CCTS tip into the well of the FCC2 to dampen the tip with optical cleaning uid.
3 Place the damp tip in the adapter until it touches the ferrule.
4 Rotate the tip clockwise 10 revolutions while applying varying pressure to create a gentle
pumping action where the tip contacts the ferrule.
5 Discard the CCTS stick after using both tips.
If using AFL ACT cleaning sticks
1 Place the ACT tip in the adapter until it touches the ferrule.
2 Ensure that stick is held straight when inserting into sleeve.
3 Apply sufcient pressure (approx. 600-700g) to ensure ferrule is a little depressed in sleeve.
4 Rotate stick clockwise 4-5 times, while ensuring direct contact with ferrule end-face is
maintained.
5 Discard the ACT stick after each use.