DX-D 100 Mobile X-ray Unit
User Manual
30
0188K EN 20201116
GUIDANCE AND MANUFACTURER’S DECLARATION - ELECTROMAGNETIC IMMUNITY
(IEC 60601-1-2:2007)
This X-ray System is intended for use in the electromagnetic environment specified below.
The customer or the user of this X-ray System should assure that it is used in such an environment.
Immunity test
IEC 60601-1-2:2007
Test Level
Compliance level Electromagnetic environment -- guidance
Conducted RF
IEC 61000--4--6
Radiated RF
IEC 61000--4--3
3Vrms
150kHzto80MHz
3V/m
80 MHz to 2.5 GHz
3Vrms
150kHzto80MHz
3V/m
80 MHz to 2.5 GHz
Portable and mobile RF communications
equipment should be used no closer to any part
of this X-ray System, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
where ’P’ is the maximum output power rating
of the transmitter i n watts (W) according to the
transmitter manufacturer and ’d’ is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site
survey
a
), should be l ess than the compliance
level in each frequency range
b
).
Interference may occur in the vicinity of
equipment marked with the following symbol:
d = 1.2 P
d = 2.3 P
, 800 MHz to 2.5 GHz
d = 1.2 P
, 80 MHz to 800 MHz
NOTE 1 -- At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 -- These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
a)
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast c annot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength i n the
location in which this X-ray System is used exceeds the applicable RF compliance l evel above, this X -ray System should be observed
to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or
relocating this X-ray System.
b)
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.