DX-D 100 Mobile X-ray Unit
User Manual
24
0188K EN 20201116
2.7.9 DISTRIBUTION OF STRAY RADIATION
Measurement conditions to determine the distribution of Stray Radiation in the
Significant Zone of Occupancy are in accordance with Standard IEC
60601-1-3:1994, IEC 60601-1-3:2008 and IEC 60601-1-3:2008+AMD1:2013.
Exposure Parameters: RAD mode, 150 kVp, 20 mAs.
Collimator opening for Field Size 18 x 18 cm, SID 100 cm.
Phantom: Rectangular water phantom of 25 x 25 x 15 cm, or a material
having a similar X-Ray attenuation coefficient.
Radiation measuring instrument: Low Radiation Dosimeter.
The results have been obtained with a configuration that is
representative of the worst case within the different configurations
of the unit.
Refer to Illustration 2-1 for position of the X-ray Unit during radiographic
examination on the Chest Unit or Front Panel, and refer to Illustration 2-2 for
position of the X-ray Unit during radiographic examination on any Patient
Support or any Table.
The following illustrations show the Distribution of Stray Radiation in each
examination position.
Note .