DX-D 100 Mobile X-ray Unit
User Manual
0188K EN 20201116
79
6.2 X-RAY TUBES
Maximum Power kW
(Refer to Identification Label)
20 kW 32 kW 40 kW 50 kW
Standard X-ray Tubes E7865X E7884X
Optional X-ray Tubes E7884X --
E7865X
Low Speed -- Rotating Anode, Focal Spots: 0.3 mm / 1.0 mm
Anode kHU / kVp: 140 kHU / 150 kVp, Target Angle: 12°
Maximum Specified Energy Input in 1 hour: 150 kVp @ 1440 mAs
Inherent Filtration of X-ray Source (Tube + Collimator): refer to Identification Label
E7884X
Low Speed -- Rotating Anode, Focal Spots: 0.6 mm / 1.2 mm
Anode kHU / kVp: 300 kHU / 150 kVp, Target Angle: 12°
Maximum Specified Energy Input in 1 hour: 150 kVp @ 3408 mAs
Inherent Filtration of X-ray Source (Tube + Collimator): refer to Identification Label