DX-D 100 Mobile X-ray Unit
User Manual
78
0188K EN 20201116
Generator Model DX- D 100
100 / 110 / 120 / 127 / 220 / 230 / 240 V~ -- Single-Phase 50 / 60 Hz
Automatic Line Compensation
¦10% V~
Connection to standard outlets with GND that complies with local regulations
Power Line Operation
The General Circuit Breaker installed in the Mobile Unit is 10 A (1P+N curve type D),
the Power Line Installation should be provided with a Differential of 30 mA Sensitivity
and with a Thermomagnetic Switch / Circuit Breaker of:
² 13 A (curve type D) or ² 20 A (curve type C) or ² 32A(curvetypeB)
Power Line Impedance must be less than the maximum indicated value:
1.2 for 110 V~, 2.5 for 230 V~
Maximum Input Power 1.5 kVA
Operation independent from
mains supply (Stand-Alone)
Standard
Battery Capacity
for the Generator
Batteries fully charged float voltage of approx. 420 Volts at nominal of approx. 382 Volts.
Charge Capacity is: 14 Ah for Lead-Crystal Batteries
The required time for the Batteries to be fully charged is approximatel y:
9 hours for Lead-Crystal Batteries
The maximum Storage Energy Capacity is: 137500 mAs @ 80 kVp
(This is the maximum energy avail able for making Exposures and supplying energy to the Generator)
The Mobil e Unit in Stand-Alone (disconnected from the mains) will be 100% discharged from full charge
in approximately: 9 hours for Lead-Crystal Batteries
Battery Capacity
for the Motors
Batteries fully charged float voltage of approx. 112 Volts at nominal of approx. 102 Volts.
Charge Capacity 9 Ah
The required time for the Batteries to be fully charged is 6 hours.
With the Batteries fully charged and disconnected from the mains,
the Mobile Unit can be in continuous movement during 4 hours (around 20 km).
If the Mobile Unit is left on in Stand-Alone (disconnected from the mains)
during 40 hours, it will be 100% discharged from full charge.
Radiation Output Accuracy
(Reproducibility related to loading factors)
C.V. (Coefficient of variation) ≤ 0.05
Maximum Symmetrical
Radiation Field
Measured at 75 kV: 200 mm in “X” axis and 260 mm in “Y” axis.
Measured at 125 kV: 200 mm in “X” axis and 260 mm in “Y” axis.
(Test performed at a distance from the Focal Spot of 1200 mm, in accordance with IEC 60806:1984).
Maximum Heat Output 260 W (1130 BTU/h)
Storage / Transport
Environmental Conditions
Temperature range of --15
o
Cto40
o
C
Relative Humidity range of 20% to 90%
Atmospheric Pressure range of 700 hPa to 1060 hPa
Operating
Environmental Conditions
Temperature range of 10
o
Cto35
o
C
(the recommended temperature for a longer life cycle of batteries is:
15
o
C
~
25
o
C for Lead-Crystal Batteries and 22
o
C
~
25
o
C for Gel Batteries)
Relative Humidity (no condensing) range of 30% to 75%
Atmospheric Pressure range of 700 hPa to 1060 hPa