DX-D 100 Mobile X-ray Unit
User Manual
60
0188K EN 20201116
Due to geometric restrictions related to the anode angle of the
X-Ray Tube, a minimum SID is required to c over the full image size
of the Detector, depending on the Collimator position:
X-Ray Tube
n
o
d
e
n
g
l
e
Detector Size
Required SID
with Collimator rotated at:
node
ngle
e
t
e
c
t
o
r
S
i
z
e
0° or ¦90° ¦45°
24X30
30X24
SID 65 cm SID 85 cm
12°
35X43
43X35
SID 90 cm SID 125 cm
43X43
S
I
9
0
c
m
S
I
1
2
5
c
m
24X30
30X24
SID 55 cm SID 65 cm
16°
35X43
43X35
SID 75 cm SID 90 cm
43X43
S
I
7
5
c
m
S
I
9
0
c
m
Note .