Install the Seal Wash Function (Quaternary Pump)
Secondary Pump Head Parts
7
Install tubing connections to the pump head.
Next Steps:
8
Install the top cover and put the module to the stack including
tubing, capillary and cable connections.
9
Route the inlet of the wash tubings into a bottle filled with a
mixture of distilled water and isopropanol (90/10) and place
the bottle in the solvent cabinet. Route the outlet of the wash
tube to the waste container.