Pump Head Maintenance (Binary and High Speed Pumps)
2
Secondary Pump Head
3
G4220A/B: Remove the front part of the pump head
including pump chamber housing with pump seal and
seal holder. If the seal wash option is installed, also
remove the support ring with wash seal and gasket.
4 G7120A, G7104A: Remove the front part of the pump
head including pump chamber housing, distance sheet,
pump seal and seal holder, and support ring with wash
seal and gasket.
5 Remove the piston from the piston housing by pushing it
to the rear, then pull it out from the rear.
6 Check the pump pistons for scratches, grooves, and
dents.
NOTE
Damaged pistons cause micro leaks and will
decrease the lifetime of the seals.
Pump chamber
housing
Seal holder
including backup
ring
Seal wash gasket
Support ring with
wash seal
EjbeX]VbWZg
]djh^c\
9^hiVcXZh]ZZi
HZVa]daYZg
HZValVh]\Vh`Zi
Hjeedgig^c\
Piston surface

Table of Contents