Pump Head Maintenance (Binary and High Speed Pumps)
2
Replacing Wash Seal and Gasket
Replacing Wash Seal and Gasket
Tools required p/n Description
01018-23702 Insert tool
Parts required p/n Description
G4220-68742 Seal Wash PM kit for 1290 Binary Pumps includes film washers, PE wash
seals and a Peristaltic pump
1 Use the steel side of the insert tool for removing the
wash seal.
2 Use the soft plastic side of the insert tool for inserting
the wash seal into the support ring. Wet the wash seal
with isopropanol or another suitable solvent before
insertion.
Wash seal
>chZgiidda
LVh]hZVa
Hjeedgig^c\
Heg^c\
HZVa