Pump Head Maintenance (Quaternary and Flexible Pumps)
3
Secondary Pump Head
Replacing the Heat Exchanger
Tools required p/n Description
Wrench, 19 mm
5023-2501 Screwdriver Torx-T10
5067-5688 Torque wrench 1 – 25 Nm with 14 mm wrench
G4220-20013 4 mm hex bit
G4220-20015 Adapter ¼ in square to hex
G4220-20041 Bit Torx 10x25 mm
Parts required p/n Description
G4220-81013 Heat Exchanger (secondary pump head only)
Preparations Remove the pump head assembly from the pump
Remove the secondary pump head from the link plate
CAUTION
Loss of small spacer fitting
Inside the secondary pump head is a small spacer fitting, which can be dropped easily
when removing the heat exchanger.
The heat exchanger does not need to be removed for pump head maintenance.