2 Pump Head Maintenance (Binary and High Speed Pumps)
Assembling the Pump Head
CAUTION
Damage to the pump head
Using a wrong torque will damage the pump head.
For handling the torque wrench, setting and applying
the right torque, consult the manual of your torque
wrench.
19 Install the pump head assembly by fixing the 4 screws
using a 4 mm hex key and a torque wrench, which are
included to the 1290 Infinity Service Kit p/n 5067-4699
set to 5 Nm. Install screws step by step, not screw by
screw.
20 Install flow connections as described in Replacing the
Pump Heads in the user manual.
5 Nm
Degasser outlet
Outlet of secondary
pump head
Inlet of primary
pump head
Purge valve
inlet
Seal wash tubing